      "status": "Ongoing",
      "city": "Milpitas",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95729",
      "recalling_firm": "Becton, Dickinson and Company, BD Bio Sciences",
      "address_1": "155 N Mccarthy Blvd",
      "address_2": "",
      "postal_code": "95035-5102",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states of Alabama, Arizona, Arkansas, California, Florida, Georgia, Illinois, Indiana, Iowa, Kentucky, Maryland, Massachusetts, Michigan, Minnesota, Nebraska, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Tennessee, Texas, Virginia & Wisconsin. The countries of China, Costa Rica, Singapore, Thailand.",
      "recall_number": "Z-0745-2025",
      "product_description": "FACSLyric 2L4C Instrument US, Catalog Number 662875, UDI 00382906628756    BD FACSLyric\" flow cytometer is intended for in vitro diagnostic for the following:  \" Immunophenotyping using up to six fluorescence detection channels and two light scatter channels using a blue (488-nm) and a red (640-nm) laser.  \" It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.",
      "product_quantity": "15 power supplies",
      "reason_for_recall": "On October 18, 2024, BD has identified through customer complaints and an increase in field failures / replacements of the power supply module (part number 650781, Power Module AC-DC 250W 12/24V M3S4S6) on BD FACSLyric flow cytometers, which has the potential for the instruments to fail to power on and/or stay on. The potential power supply failure   on BD FACSLyric Flow Cytometer may lead to risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. The patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In addition, patient specimen loss and/ or delay providing results of testing results. There may be additional risks to the laboratory staff including inhalation of fumes followed by shortness of breath or coughing if failure occurs during laboratory working hours. In addition, the patient may be asked to return for additional biospecimen collection procedure and risks associated with the procedure (bruising, pain, bleeding, etc.). In the Notification Letter sent on November 20, 2024, customers are instructed to continue their normal operation of your BD FACSLyric Flow Cytometer according to the Instructions For Use after immediately inspecting inventory for the specific catalog and power supply serial numbers listed in the notification letter and follow the instructions listed under Actions for Clinical Users",
      "recall_initiation_date": "20241101",
      "center_classification_date": "20241223",
      "report_date": "20250101",
      "code_info": "Power supply part number 650781 with serial numbers beginning with V2309, V2318, and V2330 are impacted  Catalog Number: 662875 UDI-DI code: 00382906628756 Instrument Serial Numbers/ Power Supply Serial Numbers: Z662875000119 V23300114 Z662875000117 V23300192 Z662875000112 V23300098 Z662875000123 V23300321 Z662875000105 V23090179 Z662875000104 V23090157 Z662875000107 V23090507 Z662875000102 V23090029 Z662875000103 V23090114 Z662875000106 V23090177 Z662875000108 V23180127 Z662875000110 V23300074 Z662875000111 V23090430 Z662875000109 V23300294 Z662875000116 V23300051",
      "more_code_info": ""
    }
  ]
}