      "status": "Ongoing",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95783",
      "recalling_firm": "Nova Biomedical Corporation",
      "address_1": "200 Prospect St",
      "address_2": "",
      "postal_code": "02453-3407",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.",
      "recall_number": "Z-0683-2025",
      "product_description": "StatStrip Glucose Ketone (mmol/L) Hospital Meter System. Model/Catalog Number: 63910. Prescription use blood glucose ketone meter for near-patient testing",
      "product_quantity": "1,615 units",
      "reason_for_recall": "A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.",
      "recall_initiation_date": "20241120",
      "center_classification_date": "20241220",
      "report_date": "20250101",
      "code_info": "Model No 63910.   UDI-DI (01)1038548063910.   Software Versions: v0.0.13.10 to v0.0.13.44.",
      "more_code_info": ""
    }
  ]
}