      "status": "Ongoing",
      "city": "Sparks",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95636",
      "recalling_firm": "Becton Dickinson & Co.",
      "address_1": "7 Loveton Cir",
      "address_2": "",
      "postal_code": "21152-9212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Benin, Bhutan, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Burkina Faso, Burundi, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, DOM, Dominica, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Fiji, Finland, France, French Guiana, Gambia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guam, Guatemala, Guinea, Guyana, Haiti, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Laos, Latvia, Lebanon Liberia, Libyan Arab Jamahiriya, Liechtenstein, Lithuania, Luxembourg, Macau, Madagascar Malawi, Malaysia, Maldives, Mali, Malta, Martinique, Mauritania, Mauritius, Mayotte Mexico, MKD, Moldova, Mongolia, Morocco, Mozambique, Namibia, Nepal, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Nigeria, Northern Mariana Islands, Norway, Oman, Pakistan, Panama, Papua new Guinea, Paraguay, Peru, Philippines PNG, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Rwanda, Saint Helena San Marino, Saudi Arabia, Senegal, Serbia, Seychelles, Sierra Leone, Singapore, Slovakia Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tajikistan, Tanzania, United Republic of Thailand, Timor-Leste, Tonga, Trinidad And Tobago, Tunisia, Turkey, Turkmenistan Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay Uzbekistan, Vietnam, Yemen, Zambia, Zimbabwe.",
      "recall_number": "Z-0595-2025",
      "product_description": "BACTEC 9120 System  Version or Model: 445570  Catalog Number: 445570",
      "product_quantity": "1,087 units (136 US, 951 OUS)",
      "reason_for_recall": "BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.",
      "recall_initiation_date": "20241031",
      "center_classification_date": "20241203",
      "report_date": "20241211",
      "code_info": "Catalog No: 445570; UDI-DI: 00382904455705; Serial Numbers: UB4523, UB3416, UB3975, UB3871A, UB3753, UB4823, UB3905, DB2572, UB4642, UB3414, UM1431, UB3731, UB1470, UB5555, UM1395, UB4212, UB3555, UB4159, UB4778, UB3805, UB1648, UB4562, UB4174, UB4325, UB1801, UB5086, UB1990, UB4002A, UB3103, UB3795, UB4274, UB4360, UB1611, UB4535, UB4542, UB4596, UB3338, UB4688, UB3480, UB4928, UB4929, UB1439, UB1583, UB5178, UB3122, UB3335, UB3229, UB4005, UB1682, UB3541, UB3599, UB3193, UB4184, UB1778, UB4405, UB3413, UB4271, UB4715, UB3210, UB4110, UB4112, UB3499, UB5405, UB5389, UB5442, UB5362, UB4995, UB1812, UB4427, UB4253, UB1708, UB3273, UM1273, UB4959, US1103, UB3383, UM1283, UM1269, UB5663, UB4384, UB5569, UB5470, UB5594, UB4185, UB5679, UB5527, UB5361, UB5579, UB4280, UB5535, UB4221, UB5596, UB5576, UB5588, UB5496, UB5610, UB3242, UB5537, UB5553, UB5565, UB5451, UB5505, UB5492, UB5507, UB5526, UB5691, UB5693, UB5694, UB5456, UB5609, UB5534, UB5631, UB5469, UB5541, UB3647, UB4048, UB4177, UB5415, UB5080, UB5529, UB5587, UB3398, UB5584, UB5414, UB5444, UB5474, UB5495, UB5523, UB5503, UB5431, UB3814, UB5550, UB5559, UB4661, UB5073, UB4776, UB5074, UB5059, UB5380, UB3491, UB4922, UB1862, UB4881, UB5093, UB2082, UB5079, UB4711, UB5101, UB5092, UB5108, UB5038, UB4704, UB4953, UB5026, UB5017, UB4188, UB5110, UB5041, UB5078, UB5118, UB5598, UB5111, UB5153, UB3634, UB5149, UB5005, UB5152, UB5164, UB5129, UB5050, UB5151, UB5165, UB5158, UB9654, UB5135, UB5207, UB5195, UB5171, UB5205, UB5117, UB5229, UB5241, UB5234, UB5230, UB5140, UB5167, UB5222, UB5255, UB5063, UB5199, UB5224, UB5265, UB5136, UB5283, UB5285, UB5304, UB5060, UB5289, UB5246, UB4369, UB5309, UB5302, UB5299, UB3148, UB5235, UM1331, UB1653, UB5420, UB3654, UB5329, UB4315, UB3026, UB3336, UB1501, UB3767, UB4334, UB5373, UB5791, UB5156, UB4560, UB5300, UB4519, UB5463, UB4340, UB4244, UB5472, UB3703, UB5296, UB5457, UB5317, UB3704, UB5465, UB5393, UB4356, UB3700, UB5279, UB5130, UB5297, UB4600, UB5471, UB5479, UB4545, UB4466, UB5428, UB5488, UB3851, UB5308, UB5253, UB3483, UB5298, UB4602, UB4471, UB5339, UB3812, UB4518, UB5324, UB5381, UB4504, UB5364, UB3785, UB5397, UB4539, UB4540, UB4420, UB5358, UB5433, UB3919, UB5434, UB4481, UB5425, UB5502, UB4346, UB4507, UB5371, UB5141, UB5379, UB3320, UB4610, UB5244, UB4554, UB4567, UB4620, UB4057, UB4424, UB4634, UB5218, UB4614, UB3894, UB5316, UB5483, UB5303, UB3863, UB5374, UB4115, UB5375, UB4583, UB4114, UB4392, UB5481, UB5353, UB4581, UB4006, UB4088, UB4751, UB5365, UB4666, UB1049, UB1523, UB3082, UB5466, UB5360, UB1734, UB5454, UB3062, UB2013, UB1689, UB3264, UB4571, UB5412, UB4684, UM1406, UB5581, UB1555, UB5319, UB4485, UB5589, UB4653, UB5406, UB4705, UB5396, UB1746, UB4706, UB5290, UB5432, UB4572, UB4117, UB4745, UB5543, UB4175, UB5514, UB5604, UB4730, UB3907, UB5372, UB4719, UB4681, UB4162, UB5245, UB4768, UB5388, UB4483, UB4718, UB5119, UB4139, UB5301, UB4232, UB5367, UB4723, UB5125, UB5382, UB5142, UB4747, UB3489, UB4641, UB4618, UB5061, UB4633, UB4817, UB4638, UB5418, UB4226, UB5417, UB4835, UB4643, UB4809, UB5262, UB5221, UB5422, UB5551, UB5261, UB4834, UB4756, UB4838, UB4812, UB4766, UB4810, UB4826, UB5621, UB4781, UB4861, UB4856, UB4859, UB4876, UB4854, UB5620, UB5540, UB4860, UB5538, UB4765, UB5560, UB5506, UB4755, UB5530, UB5585, UB5574, UB4902, UB4888, UB4886, UB3237, UB4870, UB4914, UB4880, UB4901, UB4906, UB4935, UB4932, UB4805, UB4917, UB4934, UB4918, UB4955, UB4905, UB4956, UB4944, UB4963, UB4916, UB4964, UB4969, UB4806, UB4979, UB4983, UB4135, UB5030, UB5013, UB4994, UB5004, UB5034, UB5020, UB4374, UB4478, UB4612, UB4651, UB4738, UB3718, UB3811, UB4463, UB3865, UB3868, UB3877, UB3878, UB3911, UB3123, UB3914, UB3915, UB5636, UB5638, UB5640, UB5642, UB5641, UB3940, UB5646, UB5650, UB5654, UB1248, UB9256, UB5347, UB5345, UM1239, UB1792, UB0031, UB3709, UB4154B, UB4740, UB4732, UB3479, UB1793, UB1750, UB1057, UB1609, UB3139, UB3853, UB3205, UB3448, UB3539, UB3779, UM1180, UM1201, UB0097, UM1345, UM2069, UB1048, UB3696, UB5102, UB5209, UB5607, UB3655, UB5561, UB5517, UB4461, UB5180, UB3948, UB3852, UB3984, UB3265, UB5036, UB4930, UB1671, UM1403, UB3952, UB5668, UB5015, UB4899, UB4129, UB4093, UB4894, UB3191, UB3595, UB4981, UB4147, UB5133, US1048, UB4260, UB4587, UB3275, UB4565, UB5268, UB5392, UB5453, UB4986, UB4475, UB5387, UB5196, UB5042, UB5443, UB5210, UB3813, UB3658, UB5223, UB4637, UB3561, UB5459, UB4201, UB4703, UB3013, UB4052, UB1977, UB5132, UB5064, UB5175, UB5247, UB4419, UB4862, UB4866, UB5557, NB3896, UB5292, UB4412, UB3553, UB4700, UB1672, UB4489, UB4491, UB5404, UB5002, UB4113, UB4487, UB3876, UB4259, UB1657, UB3170, UB1892, UB1925, UB4658, UB1716, UB5105, UB3010, UB4379, UB3409, UB5021, UB3259, UB1891, UB1668, UB1878, UB5344, UB5342, UB5512, UB5441, UB5137, UB5052, UB3721, UB2051, UB5286, UB3396, UB4181, UB2041, UB3950, UB5024, UB4394, UB5661, UB5662, UB5511, UB5563, UB5675, UB1553, UB5623, UB5508, UB5676, UB5608, UB5685, UB5447, UB5546, UB5341, UB5409, UB5438, UB5318, UB5577, UB5413, UB5625, UB5522, UB5545, UB5570, UB5627, UB5595, UB5485, UB5352, UB5462, UB5572, UB5612, UB3745, UB5593, UB5548, UB5619, UB5700, UB5586, UB5567, UB5582, UB5583, UB5439, UB5613, UB5578, UB5532, UB5562, UB5599, UB5571, UB5615, UB5558, UB5519, UB5590, UB4043, UB4220, UB5084, UB5057, UB5043, UB5053, UB5076, UB5087, UB5071, UB5099, UB5068, UB5077, UB5098, UB5066, UB5029, UB5070, UB5094, UB5045, UB5095, UB5122, UB5085, UB5022, UB4977, UB5104, UB5116, UB5124, UB5072, UB5120, UB5138, UB5145, UB5147, UB5155, UB5157, UB5169, UB5170, UB8041, UB5168, UB5146, UB5143, UB5096, UB4731, UB4904, UB5160, UB5106, UB5201, UB5197, UB5618, UB5154, UB5225, UB5243, UB5206, UB5212, UB4873, UB4631, UB5254, UB5267, UB5184, UB5211, UB5203, UB5194, UB5240, UB5219, UB5677, UB5277, UB5148, UB5242, UB5294, UB5239, UB5275, UB5131, UB5305, UB5295, UB5326, UB5573, UB5282, UB5368, UB3055, UB5385, UB3570, UB5270, UB5314, UB3042, UB5323, UB3493, UB5461, UB5429, UB4231, UB3636, UB3706, UB1934, UB3424, UB4245, UB5313, UB5359, UB5284, UB5416, UB5458, UB5260, UB5467, UB4333, UB4354, UB5435, UB4339, UB4872, UB5306, UB4342, UB5272, UB3768, UB5159, UB4355, UB5338, UB5202, UB4423, UB5516, UB5486, UB5327, UB5370, UB3800, UB4438, UB5320, UB5348, UB4205, UB3750, UB4505, UB5322, UB5307, UB5357, UB3771, UB5256, UB4410, UB4510, UB5331, UB5424, UB5489, UB4566, UB5123, UB4011, UB5408, UB4553, UB3350, UB4068, UB4486, UB4558, UB4070, UB5264, UB4083, UB5334, UB4503, UB4482, UB4498, UB5377, UB4555, UB5494, UB5177, UB5355, UB4648, UB4116, UB5500, UB4474, UB4680, UB5293, UB4720, UB5335, UB5291, UB5468, UB4773, UB5473, UB4102, UB5198, UB5455, UB5399, UB5259, UB4178, UB4685, UB4721, UB5386, UB5366, UB4105, UB4758, UB4775, UB4764, UB5491, UB5515, UB5448, UB4743, UB4759, UB5501, UB5394, UB5449, UB4464, UB5419, UB5346, UB4686, UB4163, UB5263, UB4225, UB5407, UB4763, UB5310, UB5340, UB5450, UB4767, UB5332, UB4237, UB5383, UB5333, UB4761, UB4782, UB4422, UB4871, UB5232, UB5351, UB4821, UB4671, UB5069, UB5597, UB4807, UB5547, UB4837, UB4804, UB5611, UB5464, UB5624, UB5436, UB5552, UB5549, UB4855, UB4878, UB4748, UB5518, UB5564, UB5440, UB4762, UB5498, UB4892, UB4462, UB5437, UB4885, UB4780, UB5112, UB4848, UB5065, UB5173, UB4779, UB4879, UB5533, UB4889, UB5524, UB4887, UB4852, UB4909, UB4882, UB4717, UB4948, UB4936, UB4939, UB4942, UB4853, UB4839, UB4744, UB4836, UB4891, UB4777, UB4954, UB2083, UB4884, UB4808, UB3132, UB4774, UB4965, UB4957, UB4403, UB4933, UB3680, UB4968, UB4984, UB4989, UB4679, UB4940, UB4769, UB2027, UB4198, UB4970, UB5037, UB5031, UB4982, UB5018, UB5014, UB4985, UB5044, UB4927, UB3432, UB4608, UB4632, UB4644, UB1983, UB4084, UB5602, UB5614, UB5622, UB5616, UB5634, UB3912, UB5630, UB3929, UB5645, UB5643, UB5649, UB5651, UB5653, UB5655, UB4064, UB2126, UB3304, UB3366, UB3481, UB4913, UM1179, UM1730, US1113, UB3440, UB4047, UB4261, UB4456, UB4425, UB4497, UB4820, UB4911, UB3452, UB1458, UB3482, UB6515, UB4173, UB4534, UB4662, UB1952, UM5032, UB4381, US1125, UB4363, UB4515, UB4692, UB4695, UB4365, UB4273, UB4603, UB4657, UB1629, UB4347, UB4697, UB4582, UB1673, UB4303, UB4351, UM1350, UB1697, UB4530, UB4490, UB3946A, UB4577, UB4598, UB4561, UB4604, UB4592, UB3677, UB3065, UB4296, UB4587A, UB4750, UB3318, UB4771, UB3770, UB3007, UB3992, UB3971, UB1474, UB3964, UB4050, UB3976, UB4868, UB4059, UB4101, UB1622, UB4617, UB1519, UB3004, UB1813, UB4298, UB1912, UB4179, UB4285, UB3286, UB3299, UB4118, UB4209, UB4432, UB1915, UB4670, UB4263, UB3310, UB4621, UB4301, UB4211, UB4270, UB2076, UB3023, UB3197, UB3411, UB3157, UB2026, UB2022, UB4319, UB3324, UB1893, UB3333, UB4429, UB1713, UB4233, UB3445, UB3296, UB3272, UB3225, UB2078, UB3377, UB3236, UB3332, UB4446, UB3648, UB4707, UB3551, UB3513, UB4712, UB4401, UB3303, UB3317, UB4343, UB4945, UB5249, UB5010,",
      "more_code_info": ""
    }
  ]
}