  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-14",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "arlington",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99651",
      "recalling_firm": "Wright Medical Technology, Inc.",
      "address_1": "11500 memphis arlington rd",
      "address_2": "",
      "postal_code": "38117-5423",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN, NJ, WI, CO.",
      "recall_number": "Z-3334-2026",
      "product_description": "CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System",
      "product_quantity": "34",
      "reason_for_recall": "Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.",
      "recall_initiation_date": "20260818",
      "center_classification_date": "20260930",
      "report_date": "20261007",
      "code_info": "UDI-DI: 00840420109705, Lot: 1826666, Distributed: 18-Aug-2025 through 09-Apr-2026",
      "more_code_info": ""
    }
  ]
}