  "results": [
    {
      "status": "Ongoing",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99589",
      "recalling_firm": "Tandem Diabetes Care, Inc.",
      "address_1": "12400 High Bluff Dr",
      "address_2": "",
      "postal_code": "92130-3077",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-3330-2026",
      "product_description": "Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No",
      "product_quantity": "177",
      "reason_for_recall": "A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.",
      "recall_initiation_date": "20260716",
      "center_classification_date": "20260929",
      "report_date": "20261007",
      "code_info": "Lot Code: UDI-DI: 00389152470641 Software Version: 2.13 (iOS only)",
      "more_code_info": ""
    }
  ]
}