  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-14",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sandpoint",
      "state": "ID",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99617",
      "recalling_firm": "Percussionaire Corporation",
      "address_1": "130 Mcghee Rd Ste 109",
      "address_2": "",
      "postal_code": "83864-8409",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.",
      "recall_number": "Z-3319-2026",
      "product_description": "Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065;  Phasitron 5 Breathing Circuit with Unified Connector, Product Number: P5-UC, used with In-Line Valve Instruction Manual",
      "product_quantity": "61,858",
      "reason_for_recall": "Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation",
      "recall_initiation_date": "20260818",
      "center_classification_date": "20260928",
      "report_date": "20261007",
      "code_info": "Product Number/UDI-DI:  S00065/00849436000679, 00849436000709; P5-UC/00849436000686 In-Line Valve Instruction Manual: P20174 Rev D",
      "more_code_info": ""
    }
  ]
}