      "status": "Ongoing",
      "city": "Sittingbourne",
      "state": "",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99816",
      "recalling_firm": "NOXBOX LTD",
      "address_1": "Unit 1",
      "address_2": "Eurolink Gateway; Castle Road",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the state of Georgia.",
      "recall_number": "Z-3306-2026",
      "product_description": "NOxBOXi Nitric Oxide Delivery System. Model Number: NOXBOX-I.",
      "product_quantity": "1,667 units",
      "reason_for_recall": "Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors.",
      "recall_initiation_date": "20260914",
      "center_classification_date": "20260925",
      "report_date": "20261007",
      "code_info": "Model Number: NOXBOX-I. UDI-DI: 05060541640009. Serial Numbers: NI100825, NI100833, NI100839, NI100840, NI100841, NI100847, NI100848, NI100849, NI100850, NI100851, NI100852, NI100855, NI100858, NI100859, NI100861, NI100862, NI100865, NI100866, NI100868, NI100869, NI100872, NI100873, NI100875, NI100876, NI100878, NI100879, NI100880, NI100881, NI100882, NI100883, NI100884, NI100885, NI100886, NI100887, NI100888, NI100895, NI100932, NI100935, NI100942, NI100948, NI100949, NI100953, NI100961, NI100975, NI100976, NI100978, Serial Number, NI100980, NI100983, NI100985, NI100986, NI100987, NI100989, NI100990, NI100991, NI100993, NI100994, NI100996, NI100997, NI100999, NI101001, NI101002, NI101003, NI101004, NI101005, NI101006, NI101008, NI101009, NI101010, NI101011, NI101012, NI101013, NI101014, NI101015, NI101017, NI101018, NI101019, NI101024, NI101026, NI101028, NI101029, NI101030, NI101031, NI101033, NI101034, NI101035, NI101038, NI101039, NI101041, NI101042, NI101044, NI101048, NI101049, Serial Number, NI101052, NI101057, NI101059, NI101060, NI101061, NI101063, NI101065, NI101067, NI101069, NI101071, NI101072, NI101073, NI101077, NI101079, NI101080, NI101081, NI101083, NI101084, NI101086, NI101088, NI101089, NI101090, NI101091, NI101092, NI101093, NI101094, NI101095, NI101099, NI101100, NI101101, NI101105, NI101106, NI101107, NI101108, NI101109, NI101110, NI101111, NI101112, NI101113, NI101114, NI101115, NI101116, NI101117, NI101125, NI101130, NI101131, Serial Number, NI101132, NI101133, NI101134, NI101135, NI101136, NI101137, NI101138, NI101141, NI101142, NI101144, NI101145, NI101146, NI101147, NI101148, NI101149, NI101150, NI101151, NI101153, NI101154, NI101155, NI101156, NI101158, NI101159, NI101160, NI101161, NI101162, NI101163, NI101164, NI101165, NI101167, NI101168, NI101170, NI101173, NI101174, NI101175, NI101176, NI101177, NI101178, NI101179, NI101180, NI101181, NI101182, NI101183, NI101184, NI101190, NI101191, Serial Number, NI101192, NI101199, NI101202, NI101204, NI101205, NI101207, NI101208, NI101209, NI101211, NI101212, NI101213, NI101214, NI101215, NI101216, NI101217, NI101218, NI101219, NI101220, NI101221, NI101222, NI101223, NI101224, NI101225, NI101227, NI101228, NI101229, NI101230, NI101231, NI101232, NI101233, NI101234, NI101236, NI101237, NI101241, NI101243, NI101244, NI101245, NI101246, NI101249, NI101250, NI101251, NI101252, NI101253, NI101254, NI101255, NI101256, Serial Number, NI101257, NI101258, NI101259, NI101263, NI101264, NI101265, NI101266, NI101268, NI101270, NI101271, NI101273, NI101278, NI101282, NI101284, NI101285, NI101286, NI101287, NI101288, NI101289, NI101292, NI101294, NI101295, NI101297, NI101298, NI101299, NI101300, NI101301, NI101303, NI101305, NI101307, NI101308, NI101309, NI101313, NI101314, NI101317, NI101321, NI101322, NI101323, NI101324, NI101326, NI101327, NI101328, NI101329, NI101330, NI101331, NI101333, Serial Number, NI101335, NI101336, NI101337, NI101338, NI101340, NI101342, NI101344, NI101346, NI101347, NI101349, NI101350, NI101351, NI101353, NI101354, NI101355, NI101356, NI101357, NI101359, NI101360, NI101361, NI101368, NI101372, NI101373, NI101376, NI101377, NI101379, NI101380, NI101381, NI101384, NI101386, NI101387, NI101388, NI101390, NI101392, NI101393, NI101394, NI101395, NI101396, NI101402, NI101404, NI101405, NI101406, NI101407, NI101408, NI101409, NI101410, Serial Number, NI101411, NI101412, NI101413, NI101414, NI101415, NI101416, NI101423, NI101425, NI101426, NI101428, NI101429, NI101430, NI101431, NI101432, NI101433, NI101434, NI101435, NI101436, NI101437, NI101439, NI101441, NI101442, NI101443, NI101444, NI101446, NI101448, NI101449, NI101451, NI101452, NI101454, NI101456, NI101457, NI101459, NI101460, NI101461, NI101462, NI101463, NI101464, NI101467, NI101468, NI101469, NI101472, NI101473, NI101474, NI101475, NI101478, Serial Number, NI101479, NI101482, NI101483, NI101484, NI101485, NI101486, NI101487, NI101489, NI101490, NI101491, NI101492, NI101496, NI101505, NI101506, NI101508, NI101509, NI101512, NI101514, NI101515, NI101516, NI101520, NI101521, NI101522, NI101524, NI101525, NI101526, NI101527, NI101529, NI101530, NI101532, NI101533, NI101540, NI101550, NI101551, NI101552, NI101553, NI101554, NI101557, NI101560, NI101563, NI101565, NI101566, NI101569, NI101570, NI101572, NI101573, Serial Number, NI101574, NI101575, NI101576, NI101577, NI101578, NI101579, NI101581, NI101584, NI101585, NI101588, NI101590, NI101591, NI101592, NI101593, NI101594, NI101595, NI101596, NI101597, NI101598, NI101599, NI101600, NI101601, NI101602, NI101604, NI101605, NI101606, NI101607, NI101608, NI101609, NI101610, NI101611, NI101612, NI101613, NI101614, NI101615, NI101616, NI101617, NI101620, NI101622, NI101623, NI101624, NI101625, NI101626, NI101627, NI101628, NI101629, Serial Number, NI101635, NI101636, NI101638, NI101639, NI101640, NI101641, NI101643, NI101644, NI101645, NI101651, NI101654, NI101655, NI101656, NI101657, NI101658, NI101659, NI101660, NI101662, NI101664, NI101665, NI101666, NI101667, NI101668, NI101669, NI101670, NI101671, NI101672, NI101674, NI101675, NI101676, NI101677, NI101678, NI101679, NI101680, NI101681, NI101682, NI101683, NI101684, NI101685, NI101686, NI101687, NI101688, NI101689, NI101690, NI101691, NI101692, Serial Number, NI101693, NI101694, NI101695, NI101696, NI101697, NI101699, NI101700, NI101703, NI101704, NI101705, NI101706, NI101707, NI101708, NI101709, NI101710, NI101711, NI101712, NI101713, NI101714, NI101715, NI101716, NI101717, NI101718, NI101719, NI101720, NI101721, NI101722, NI101723, NI101724, NI101726, NI101727, NI101729, NI101731, NI101732, NI101733, NI101735, NI101736, NI101737, NI101739, NI101740, NI101742, NI101743, NI101744, NI101746, NI101747, NI101748, Serial Number, NI101749, NI101750, NI101751, NI101753, NI101756, NI101759, NI101760, NI101761, NI101763, NI101765, NI101766, NI101767, NI101768, NI101769, NI101770, NI101771, NI101775, NI101776, NI101777, NI101778, NI101781, NI101782, NI101783, NI101784, NI101787, NI101791, NI101793, NI101794, NI101795, NI101796, NI101797, NI101798, NI101799, NI101804, NI101805, NI101806, NI101807, NI101808, NI101809, NI101810, NI101811, NI101812, NI101813, NI101814, NI101815, NI101816, Serial Number, NI101817, NI101818, NI101819, NI101820, NI101821, NI101822, NI101823, NI101824, NI101825, NI101826, NI101828, NI101829, NI101831, NI101832, NI101834, NI101835, NI101836, NI101837, NI101838, NI101840, NI101841, NI101842, NI101843, NI101844, NI101845, NI101846, NI101848, NI101849, NI101851, NI101854, NI101857, NI101859, NI101861, NI101862, NI101864, NI101865, NI101867, NI101868, NI101869, NI101872, NI101873, NI101874, NI101875, NI101876, NI101878, NI101879, Serial Number, NI101880, NI101881, NI101882, NI101883, NI101884, NI101885, NI101886, NI101887, NI101889, NI101890, NI101891, NI101892, NI101895, NI101896, NI101898, NI101899, NI101901, NI101902, NI101903, NI101904, NI101905, NI101906, NI101907, NI101908, NI101909, NI101910, NI101912, NI101913, NI101914, NI101916, NI101919, NI101920, NI101922, NI101924, NI101925, NI101926, NI101928, NI101936, NI101937, NI101939, NI101940, NI101941, NI101942, NI101943, NI101944, NI101945, Serial Number, NI101946, NI101947, NI101948, NI101952, NI101953, NI101954, NI101955, NI101956, NI101957, NI101958, NI101959, NI101960, NI101961, NI101962, NI101963, NI101964, NI101965, NI101966, NI101967, NI101968, NI101970, NI101971, NI101972, NI101973, NI101974, NI101975, NI101976, NI101977, NI101978, NI101979, NI101981, NI101982, NI101983, NI101984, NI101985, NI101986, NI101987, NI101988, NI101989, NI101990, NI101991, NI101992, NI101993, NI101994, NI101995, NI101996, Serial Number, NI101997, NI101998, NI101999, NI102000, NI102001, NI102002, NI102003, NI102004, NI102006, NI102007, NI102008, NI102009, NI102010, NI102011, NI102012, NI102013, NI102014, NI102015, NI102017, NI102018, NI102019, NI102020, NI102021, NI102023, NI102024, NI102025, NI102026, NI102027, NI102028, NI102031, NI102032, NI102033, NI102035, NI102036, NI102037, NI102038, NI102040, NI102043, NI102044, NI102045, NI102046, NI102047, NI102048, NI102049, NI102051, NI102052, Serial Number, NI102053, NI102055, NI102056, NI102057, NI102059, NI102060, NI102061, NI102062, NI102063, NI102065, NI102067, NI102068, NI102069, NI102070, NI102071, NI102072, NI102074, NI102075, NI102076, NI102077, NI102078, NI102079, NI102081, NI102082, NI102083, NI102086, NI102089, NI102090, NI102091, NI102092, NI102093, NI102094, NI102095, NI102096, NI102097, NI102098, NI102099, NI102102, NI102103, NI102104, NI102105, NI102106, NI102107, NI102108, NI102109, NI102110, Serial Number, NI102111, NI102113, NI102114, NI102115, NI102116, NI102117, NI102118, NI102119, NI102120, NI102121, NI102123, NI102124, NI102125, NI102126, NI102127, NI102128, NI102129, NI102131, NI102132, NI102133, NI102134, NI102135, NI102136, NI102137, NI102138, NI102139, NI102140, NI102141, NI102142, NI102144, NI102145, NI102146, NI102147, NI102148, NI102149, NI102150, NI102152, NI102155, NI102160, NI102161, NI102163, NI102164, NI102165, NI102166, NI102167, NI102168, Serial Number, NI102169, NI102170, NI102171, NI102172, NI102173, NI102174, NI102175, NI102176, NI102177, NI102180, NI102182, NI102183, NI102184, NI102186, NI102188, NI102189, NI102190, NI102191, NI102192, NI102193, NI102194, NI102195, NI102196, NI102197, NI102198, NI102199, NI102200, NI102205, NI102206, NI102208, NI102209, NI102212, NI102213, NI102214, NI102215, NI102216, NI102219, NI102221, NI102222, NI102226, NI102227, NI102228, NI102230, NI102231, NI102232, NI102233, Serial Number, NI102234, NI102235, NI102237, NI102238, NI102239, NI102240, NI102241, NI102244, NI102245, NI102246, NI102247, NI102249, NI102250, NI102253, NI102254, NI102255, NI102256, NI102259, NI102263, NI102265, NI102266, NI102268, NI102269, NI102270, NI102271, NI102272, NI102273, NI102274, NI102275, NI102276, NI102277, NI102278, NI102279, NI102280, NI102281, NI102282, NI102283, NI102284, NI102285, NI102286, NI102287, NI102288, NI102290, NI102291, NI102292, NI102293, Serial Number, NI102295, NI102296, NI102297, NI102298, NI102299, NI102300, NI102304, NI102305, NI102306, NI102307, NI102308, NI102309, NI102310, NI102311, NI102327, NI102329, NI102330, NI102331, NI102332, NI102338, NI102339, NI102340, NI102341, NI102342, NI102343, NI102345, NI102346, NI102347, NI102349, NI102350, NI102351, NI102355, NI102356, NI102357, NI102358, NI102359, NI102360, NI102361, NI102362, NI102363, NI102364, NI102366, NI102377, NI102380, NI102381, NI102383, Serial Number, NI102384, NI102385, NI102387, NI102389, NI102390, NI102391, NI102392, NI102393, NI102394, NI102395, NI102396, NI102397, NI102398, NI102399, NI102400, NI102402, NI102403, NI102404, NI102405, NI102406, NI102407, NI102408, NI102410, NI102411, NI102413, NI102414, NI102415, NI102417, NI102418, NI102422, NI102423, NI102424, NI102425, NI102426, NI102427, NI102428, NI102429, NI102431, NI102432, NI102433, NI102434, NI102435, NI102436, NI102437, NI102438, NI102444, Serial Number, NI102445, NI102447, NI102463, NI102466, NI102467, NI102472, NI102473, NI102474, NI102475, NI102476, NI102477, NI102478, NI102479, NI102480, NI102481, NI102482, NI102483, NI102484, NI102485, NI102486, NI102487, NI102488, NI102489, NI102490, NI102492, NI102494, NI102495, NI102496, NI102497, NI102498, NI102499, NI102500, NI102501, NI102502, NI102503, NI102504, NI102506, NI102507, NI102509, NI102510, NI102511, NI102512, NI102513, NI102517, NI102518, NI102519, Serial Number, NI102520, NI102522, NI102525, NI102528, NI102530, NI102531, NI102535, NI102536, NI102537, NI102538, NI102539, NI102540, NI102541, NI102542, NI102543, NI102544, NI102545, NI102546, NI102547, NI102548, NI102550, NI102551, NI102552, NI102553, NI102554, NI102555, NI102556, NI102557, NI102558, NI102560, NI102561, NI102562, NI102563, NI102564, NI102565, NI102566, NI102567, NI102568, NI102569, NI102570, NI102571, NI102574, NI102575, NI102576, NI102577, NI102578, Serial Number, NI102579, NI102580, NI102581, NI102582, NI102583, NI102584, NI102585, NI102586, NI102587, NI102588, NI102590, NI102591, NI102592, NI102593, NI102594, NI102595, NI102596, NI102597, NI102598, NI102599, NI102600, NI102601, NI102602, NI102604, NI102605, NI102606, NI102607, NI102608, NI102610, NI102611, NI102613, NI102614, NI102615, NI102616, NI102617, NI102618, NI102619, NI102620, NI102621, NI102622, NI102623, NI102624, NI102625, NI102627, NI102628, NI102629, Serial Number, NI102630, NI102631, NI102632, NI102634, NI102635, NI102637, NI102638, NI102639, NI102640, NI102641, NI102642, NI102643, NI102645, NI102647, NI102648, NI102649, NI102651, NI102652, NI102653, NI102655, NI102656, NI102657, NI102658, NI102659, NI102660, NI102661, NI102662, NI102663, NI102664, NI102665, NI102666, NI102667, NI102668, NI102669, NI102670, NI102671, NI102672, NI102673, NI102674, NI102675, NI102676, NI102680, NI102681, NI102683, NI102684, NI102685, Serial Number, NI102686, NI102687, NI102688, NI102689, NI102691, NI102692, NI102693, NI102695, NI102696, NI102697, NI102698, NI102699, NI102700, NI102701, NI102702, NI102703, NI102704, NI102705, NI102706, NI102707, NI102708, NI102709, NI102710, NI102711, NI102713, NI102714, NI102715, NI102717, NI102720, NI102722, NI102724, NI102725, NI102726, NI102727, NI102728, NI102729, NI102730, NI102731, NI102732, NI102733, NI102734, NI102735, NI102736, NI102737, NI102738, NI102739, Serial Number, NI102740, NI102741, NI102742, NI102743, NI102744, NI102745, NI102746, NI102747, NI102748, NI102751, NI102752, NI102753, NI102755, NI102756, NI102757, NI102758, NI102759, NI102760, NI102762, NI102763, NI102764, NI102765, NI102766, NI102767, NI102768, NI102769, NI102770, NI102771, NI102772, NI102773, NI102774, NI102776, NI102777, NI102778, NI102781, NI102783, NI102784, NI102785, NI102786, NI102789, NI102790, NI102791, NI102792, NI102793, NI102794, NI102795, Serial Number, NI102798, NI102799, NI102800, NI102801, NI102802, NI102803, NI102804, NI102805, NI102806, NI102807, NI102808, NI102809, NI102810, NI102811, NI102812, NI102813, NI102814, NI102815, NI102816, NI102817, NI102818, NI102819, NI102821, NI102823, NI102824, NI102825, NI102828, NI102830, NI102832, NI102833, NI102834, NI102836, NI102837, NI102838, NI102839, NI102840, NI102841, NI102842, NI102843, NI102844, NI102845, NI102846, NI102847, NI102848, NI102849, NI102850, Serial Number, NI102851, NI102852, NI102853, NI102854, NI102855, NI102856, NI102858, NI102859, NI102860, NI102861, NI102862, NI102863, NI102865, NI102866, NI102868, NI102870, NI102871, NI102872, NI102873, NI102874, NI102876, NI102877, NI102878, NI102879, NI102880, NI102881, NI102882, NI102883, NI102884, NI102885, NI102886, NI102887, NI102888, NI102889, NI102890, NI102891, NI102892, NI102893, NI102894, NI102897, NI102898, NI102899, NI102900, NI102901, NI102902, NI102904, Serial Number, NI102905, NI102906, NI102907, NI102908, NI102910, NI102911, NI102913, NI102914, NI102915, NI102916, NI102917, NI102918, NI102919, NI102920, NI102921, NI102922, NI102923, NI102926, NI102927, NI102928, NI102929, NI102930, NI102931, NI102932, NI102933, NI102934, NI102935, NI102936, NI102937, NI102939, NI102942, NI102944, NI102947, NI102950, NI102955, NI102957, NI102958, NI102961, NI102962, NI102963, NI102964, NI102966, NI102967, NI102968, NI102970, NI102973, Serial Number, NI102977, NI102979, NI102986, NI102987, NI102989, NI102990, NI102991, NI102992, NI102996, NI102997, NI102998, NI102999, NI103001, NI103002, NI103003, NI103007, NI103011, NI103012, NI103014, NI103016, NI103017, NI103018, NI103019, NI103020, NI103021, NI103022, NI103023, NI103024, NI103074, NI103079, NI103082, NI103083, NI103084, NI103085, NI103087, NI103088, NI103089, NI103091, NI103093, NI103094, NI103095, NI103104, NI103116, NI103117, NI103120, NI103122, Serial Number, NI103123, NI103126, NI103127, NI103131, NI103136, NI103139, NI103144, NI103145, NI103146, NI103147, NI103148, NI103151, NI103152, NI103153, NI103155, NI103157, NI103159, NI103168, NI103170, NI103181, NI103182, NI103183, NI103189, NI103196, NI103203, NI103204, NI103208, NI103215, NI103220.   Includes NOxBOXi with Oxygen sensor lots manufactured before April 13, 2026",
      "more_code_info": ""
    }
  ]
}