  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99779",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "International distribution to the country of Japan.",
      "recall_number": "Z-3299-2026",
      "product_description": "Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.",
      "product_quantity": "72 units",
      "reason_for_recall": "There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.",
      "recall_initiation_date": "20260824",
      "center_classification_date": "20260925",
      "report_date": "20261007",
      "code_info": "Model Number: M3-30A. UDI-DI: 00821925005679. All Serial Numbers.",
      "more_code_info": ""
    }
  ]
}