  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Marlborough",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95124",
      "recalling_firm": "Boston Scientific Corporation",
      "address_1": "100 Boston Scientific Way",
      "address_2": "",
      "postal_code": "01752-1234",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of APAC, Canada, EMEA, LATAM.",
      "recall_number": "Z-3286-2024",
      "product_description": "WALLFLEX PC ESOPH STENT 23/28MM X 155MM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is Intended for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas    Material Number (UPN): M00516950",
      "product_quantity": "199 units",
      "reason_for_recall": "WallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.",
      "recall_initiation_date": "20240801",
      "center_classification_date": "20240926",
      "report_date": "20241002",
      "code_info": "GTIN: 08714729765295 Lot Numbers:  32764205, 32764206, 32827703 32837016, 32837017, 32850357 32869530, 32876613, 32891019 32900187, 32938379, 33145977 33160189, 33167134, 33242627",
      "more_code_info": ""
    }
  ]
}