  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sparks",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95145",
      "recalling_firm": "Becton Dickinson & Co.",
      "address_1": "7 Loveton Cir",
      "address_2": "",
      "postal_code": "21152-9212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Burundi, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Dominican Republic Ecuador, Egypt, El Salvador, Ethiopia, France, French Guiana, Georgia, Germany, Ghana, Greece, Guatemala, Guinea, Honduras, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Malaysia, Mali, Mauritania, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Sierra Leone, Singapore,  South Africa, South Korea, Spain, Sri Lanka, Switzerland, Syrian Arab, Republic Taiwan, Tanzania, United Republic of Thailand, Timor-Leste, Trinidad And Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, United States, Uruguay, Uzbekistan, Vietnam, Zambia, Zimbabwe.",
      "recall_number": "Z-3247-2024",
      "product_description": "BD Phoenix 100-Intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria. The BD Phoenix System provides rapid results for most aerobic and facultative anaerobic Grampositive bacteria.  Catalog Number: 448100",
      "product_quantity": "764 (EA)",
      "reason_for_recall": "Escherichia coli (E. coli) misidentification for clinical samples that were tested on BD Phoenix\" M50 instrument and BD Phoenix\" 100, may lead to misdiagnosis and inappropriate treatment of infections caused by E. coli,",
      "recall_initiation_date": "20240801",
      "center_classification_date": "20240924",
      "report_date": "20241002",
      "code_info": "UDI: 00382904481001 All serial numbers",
      "more_code_info": ""
    }
  ]
}