  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-30",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99729",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of MN and SD.",
      "recall_number": "Z-3230-2026",
      "product_description": "Medline convenience kits containing Healthmark Instrument Channel Brushes.   SINUS PACK DYNJ48258D and SINUS PACK DYNJ48258F.",
      "product_quantity": "90 kits (US)",
      "reason_for_recall": "Kits contain Healthmark Instrument Channel Brushes, which were recalled by the manufacturer due to the potential for a sharp edge/burr on the end of the brush. Potential hazards associated include cut injury/abrasion and infection.",
      "recall_initiation_date": "20260810",
      "center_classification_date": "20260918",
      "report_date": "20260930",
      "code_info": "DYNJ48258F    UDI-DI 10198459038334    Lots 24HBR684   25CBS234   25GBH438   25JBC863   26FBU052",
      "more_code_info": ""
    }
  ]
}