      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95289",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark,  Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel,  Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands,  Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore,  Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.",
      "recall_number": "Z-3205-2024",
      "product_description": "SmartPath to Ingenia Elition X - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation.     (REF) Numbers:   782118",
      "product_quantity": "2 units",
      "reason_for_recall": "Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated.  movement.",
      "recall_initiation_date": "20240903",
      "center_classification_date": "20240920",
      "report_date": "20241002",
      "code_info": "UDI-DI: (01)00884838099746(21) Serial Numbers: 85661 85676",
      "more_code_info": ""
    }
  ]
}