  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Best",
      "state": "",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95256",
      "recalling_firm": "PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.",
      "address_1": "Veenpluis 4-6",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide, Canada and Worldwide: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Brunei Darussalam Bulgaria Cambodia Cayman Islands Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cuba Curacao Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Fiji Finland France French Polynesia Georgia Germany Ghana Gibraltar Greece Guadeloupe Guinea Honduras Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Jamaica Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Kyrgyzstan Lao People's Democratic Republic Latvia Lebanon Libya Lithuania Macao Macedonia Malaysia Maldives Malta Martinique Mauritania Mauritius Mexico Mongolia Morocco Myanmar Nepal Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pakistan Palestine, State of Panama Papua New Guinea Paraguay Peru Philippines Poland Portugal Qatar R¿union Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Trinidad and Tobago Tunisia Turkey Turkmenistan Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Venezuela Viet Nam Virgin Islands (U.S.) Yemen",
      "recall_number": "Z-3168-2024",
      "product_description": "System Name: 708038 MultiDiagnost Eleva with Flat Detector 708037 MultiDiagnost Eleva with Flat Detector 708036 MultiDiagnost Eleva 708034 MultiDiagnost Eleva with Flat Detector 708032 MultiDiagnost Eleva",
      "product_quantity": "109 units",
      "reason_for_recall": "Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.",
      "recall_initiation_date": "20240822",
      "center_classification_date": "20240918",
      "report_date": "20240925",
      "code_info": "Model               Product                              UDI-DI:  N/A 708038   MultiDiagnost Eleva with Flat Detector       708037   MultiDiagnost Eleva with Flat Detector       708036   MultiDiagnost Eleva       708034   MultiDiagnost Eleva with Flat Detector       708032   MultiDiagnost Eleva        All serial numbers",
      "more_code_info": ""
    }
  ]
}