  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Inning",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95233",
      "recalling_firm": "Btt Health Gmbh",
      "address_1": "Billerberg 7",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the states of Florida.",
      "recall_number": "Z-3162-2024",
      "product_description": "Marodyne-LiV, part number 3045-133, Low intensity vibration exerciser supporting osteoporosis treatment",
      "product_quantity": "62 units",
      "reason_for_recall": "Some power supply units in Austria and Germany had a voltage drop in the millisecond range, which led to a cancellation of the treatment and subsequent restart of the medical device. The treatment is only restarted by a step-off/step-on (step-on for activation).",
      "recall_initiation_date": "20240801",
      "center_classification_date": "20240917",
      "report_date": "20240925",
      "code_info": "serial numbers:  010426065891001421S05327911240506, 010426065891001421S05328011240506, 010426065891001421S05328111240506, 010426065891001421S05328211240506, 010426065891001421S05328311240506, 010426065891001421S05328411240506, 010426065891001421S05328511240506, 010426065891001421S05328611240506, 010426065891001421S05328711240506, 010426065891001421S05328811240506, 010426065891001421S05328911240506, 010426065891001421S05329011240506, 010426065891001421S05321911240430, 010426065891001421S05322011240430, 010426065891001421S05322111240430, 010426065891001421S05322211240430, 010426065891001421S05322311240430, 010426065891001421S05322411240430, 010426065891001421S05322511240430, 010426065891001421S05322611240430, 010426065891000721S05336411240521, 010426065891000721S05336511240521, 010426065891000721S05336611240521, 010426065891000721S05336711240521, 010426065891000721S05336811240521, 010426065891000721S05336911240521, 010426065891000721S05337011240521, 010426065891000721S05337111240521, 010426065891000721S05337211240521, 010426065891000721S05337311240521, 010426065891000721S05337411240521, 010426065891000721S05337511240521, 010426065891000721S05337611240521, 010426065891000721S05337711240521, 010426065891000721S05337811240521, 010426065891000721S05337911240521, 010426065891000721S05338011240521, 010426065891000721S05338111240521, 010426065891000721S05338211240521, 010426065891000721S05338311240521, 010426065891000721S05338411240521, 010426065891000721S05338511240521, 010426065891000721S05338611240521, 010426065891000721S05338711240521, 010426065891000721S05338811240521, 010426065891000721S05338911240521, 010426065891000721S05339011240521, 010426065891000721S05339111240521, 010426065891000721S05339211240521, 010426065891000721S05339311240521, 010426065891000721S05339411240521, 010426065891000721S05339511240521, 010426065891000721S05339611240521, 010426065891000721S05339711240521, 010426065891000721S05339811240521, 010426065891000721S05339911240521, 010426065891000721S05340011240521, 010426065891000721S05340111240521, 010426065891000721S05340211240521, 010426065891000721S05340311240521, 010426065891000721S05340411240521, 010426065891000721S05340511240521",
      "more_code_info": ""
    }
  ]
}