{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-23",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "San Clemente",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99058",
      "recalling_firm": "ICU Medical, Inc.",
      "address_1": "951 Calle Amanecer",
      "address_2": "N/A",
      "postal_code": "92673-6212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Press Release",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and the countries of PK, PH, MY, HK, IN, DE, CR.",
      "recall_number": "Z-3157-2026",
      "product_description": "IV Non-Dedicated Gravity Administration Sets with drip chambers",
      "product_quantity": "2,450",
      "reason_for_recall": "Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.",
      "recall_initiation_date": "20260512",
      "center_classification_date": "20260911",
      "report_date": "20260923",
      "code_info": "All products manufactured on or before 7/4/2026. UDI-DI: 10887709079363 10887709079370 10887709079417 10887709079431 10887709079448 10887709079455 10887709097473 10887709079493 10887709079509 10887709079547",
      "more_code_info": ""
    }
  ]
}