      "status": "Ongoing",
      "city": "Allentown",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95153",
      "recalling_firm": "B. Braun Medical, Inc.",
      "address_1": "901 Marcon Blvd",
      "address_2": "",
      "postal_code": "18109-9512",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-3144-2024",
      "product_description": "AET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock, Model/Catalog Number: 472036",
      "product_quantity": "19050 units",
      "reason_for_recall": "Potential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process.",
      "recall_initiation_date": "20240808",
      "center_classification_date": "20240913",
      "report_date": "20240925",
      "code_info": "Model No 472036; UDI-DI 04046964184659; Lots 0061898815, 0061905158, 00VL872690, 00VL882317 and 00VL882319",
      "more_code_info": ""
    }
  ]
}