{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-30",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Trumbull",
      "state": "CT",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99643",
      "recalling_firm": "CooperSurgical, Inc.",
      "address_1": "95 Corporate Dr",
      "address_2": "",
      "postal_code": "06611-1350",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.",
      "recall_number": "Z-3140-2026",
      "product_description": "INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.",
      "product_quantity": "149 boxes (745 units)",
      "reason_for_recall": "Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.",
      "recall_initiation_date": "20260811",
      "center_classification_date": "20260918",
      "report_date": "20260930",
      "code_info": "1. Model Number: 44-0712. UDI-DI: (01)60888937009516. Lot Code: 550031624.   2. Model Number: 44-0707. UDI-DI: (01)20888937009488. Lot Code: 550021057.   3.  Model Number: 44-0709. UDI-DI: (01)20888937009495. Lot Code: 550031623.   4.  Model Number: 44-0715. UDI-DI: (01)20888937009525. Lot Code: 550031625.",
      "more_code_info": ""
    }
  ]
}