  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95150",
      "recalling_firm": "Datascope Corp.",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "International distribution to the countries of UNITED ARAB EMIRATES, ARGENTINA , AUSTRIA, AUSTRALIA, BELGIUM, BULGARIA, BAHRAIN, BRAZIL, CANADA, SWITZERLAND, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, GERMANY, DENMARK, ECUADOR, EGYPT, SPAIN, FINLAND, FRANCE, UNITED KINGDOM, INDONESIA, IRELAND, INDIA, IRAQ, IRAN, ITALY, JORDAN, JAPAN, SOUTH KOREA, KUWAIT, LEBANON, LITHUANIA, LIBYA, MOROCCO, MYANMAR, MEXICO, MALAYSIA, NETHERLANDS, NORWAY, NEW ZEALAND, OMAN, PHILIPPINES, POLAND, PALESTINE, PORTUGAL, SERBIA, RUSSIAN FEDERATION, SAUDI ARABIA, SWEDEN, SINGAPORE, SLOVAKIA, THAILAND, TURKEY, TAIWAN, UNITED KINGDOM, VIETNAM, YEMEN.",
      "recall_number": "Z-3135-2024",
      "product_description": "CARDIOSAVE RESCUE INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.",
      "product_quantity": "5,475 units",
      "reason_for_recall": "Firm has developed a software correction (Revision D.00) outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-Aortic Balloon Pump devices.",
      "recall_initiation_date": "20240808",
      "center_classification_date": "20240920",
      "report_date": "20241002",
      "code_info": "Product No. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85; UDI: 10607567112312, 10607567108407, 10607567113449; Lot No. ALL LOTS.",
      "more_code_info": ""
    }
  ]
}