  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95128",
      "recalling_firm": "MEDLINE INDUSTRIES, LP - Northfield",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Panama.",
      "recall_number": "Z-3107-2024",
      "product_description": "Medline convenience kits containing semi-rigid suction liners labeled as follows:    a) OPEN HEART KIT,ALTA BATES HOSP, Pack Number DYK1002174I;   b) ENDO KIT, Pack Number DYKE1558B;   c) ENDO KIT, Pack Number DYKE1804A;   d) VASCULAR-LF, Pack Number DYNJ900714C;   e) MAJOR CARDIOVASC ADULT, Pack Number DYNJ905472B;   f) AV FISTULA, Pack Number DYNJ908096A",
      "product_quantity": "1,646 units",
      "reason_for_recall": "Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.",
      "recall_initiation_date": "20240627",
      "center_classification_date": "20240911",
      "report_date": "20240918",
      "code_info": "a) DYK1002174I:  UDI/DI case 40195327332403, UDI/DI each 10195327332402; Lot Numbers:  24CLA959;  b) DYKE1558B:  UDI/DI case 40195327258512, UDI/DI each 10195327258511; Lot Numbers:  24DMA783, 24BMG878, 23LMG790;  c) DYKE1804A:  UDI/DI case 40195327023035, UDI/DI each 10195327023034; Lot Numbers:  24BMI437, 24AME311, 23LMC958;  d) DYNJ900714C:  UDI/DI case 40193489951364, UDI/DI each 10193489951363; Lot Numbers:  24FBC609;  e) DYNJ905472B:  UDI/DI case 40195327200573, UDI/DI each 10195327200572; Lot Numbers:  24BBS291;  f) DYNJ908096A:  UDI/DI case 40195327507061, UDI/DI each 10195327507060; Lot Numbers:  24ABX948, 24ABB080",
      "more_code_info": ""
    }
  ]
}