{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "East Walpole",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95118",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "333 Coney St",
      "address_2": "",
      "postal_code": "02032-1516",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution",
      "recall_number": "Z-3100-2024",
      "product_description": "Atellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers  SMN: 10995490",
      "product_quantity": "",
      "reason_for_recall": "Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample  results above a concentration of 110 U/mL.",
      "recall_initiation_date": "20240726",
      "center_classification_date": "20240911",
      "report_date": "20240918",
      "code_info": "UDI-DI: (01)00630414598178(10)55974535(17)20240914  Kit Lot: 55974535, 55975535",
      "more_code_info": ""
    }
  ]
}