      "status": "Ongoing",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95178",
      "recalling_firm": "Smith & Nephew Inc.",
      "address_1": "150 Minuteman Rd",
      "address_2": "",
      "postal_code": "01810-1031",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Chile, China, France, Germany, Great Britain, Greece, Hong Kong, India, Italy, Japan, Lithuania, Malaysia, Malta, New Zealand, Poland, Portugal, Puerto Rico, Romania, Singapore, Slovenia, South Africa, Spain, Switzerland, United Arab Emirates/Dubai (UAE).",
      "recall_number": "Z-3096-2024",
      "product_description": "ROTATION MEDICAL TENDON STAPLES (8) (broad indication)-Intended for fixation of prosthetic material to soft tissues.  Product Number: 2504-1",
      "product_quantity": "US: 6305 units;  OUS : 160 units",
      "reason_for_recall": "Product packaging process may result in an improper or incomplete seal of the outer foil pouch surrounding the inner Tyvek pouch, compromising product sterility",
      "recall_initiation_date": "20240812",
      "center_classification_date": "20240910",
      "report_date": "20240918",
      "code_info": "UDI-DI: 00854501006067 Batch Number 51180055 51180056 51180057 51180058 51180059 51180060 51180061 51180062 51180063 51180064 51180065 51184280 51184281 51184282 51184283 51184284 51184285 51184286 51184287 51184288 51184289 51184290 51185286 51185287 51185288 51185289 Batch Number 51185290 51185291 51185292 51185293 51185294 51185295 51185296 51189247 51189249 51189250 51189251 51189252 51189254 51189255 51189256 51189257 51189258 51189259 51193811 51193816 51193818 51193820",
      "more_code_info": ""
    }
  ]
}