  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-02",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Chaska",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99596",
      "recalling_firm": "IMRIS Imaging Inc",
      "address_1": "1230 Chaska Creek Way Ste 100",
      "address_2": "N/A",
      "postal_code": "55318-2707",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.",
      "recall_number": "Z-3034-2026",
      "product_description": "ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600",
      "product_quantity": "13 units",
      "reason_for_recall": "The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.",
      "recall_initiation_date": "20260806",
      "center_classification_date": "20260826",
      "report_date": "20260902",
      "code_info": "114093-000 - UDI-DI: 00857534006356, Serial Numbers: 10005695, 10002998, 10002997, 10005456, 10005612, 10004372, 10005613, 10002841, 10011119;  114093-600 - UDI-DI: 00857534006462, Serial Numbers: 10002996, 10004515, 10007411, 10007376"
    }
  ]
}