  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86273",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of CA, FL and NJ. The countries of Algeria, Egypt,  Philippines and Taiwan.",
      "recall_number": "Z-3028-2020",
      "product_description": "UIBC; Catalog Number: OSR61205; UDI (01) 15099590011925; IVD; Rx Only; WARNING H316 H317 H351 H373 H411 - Product Usage: System reagent for the quantitative determination of Unsaturated Iron Binding Capacity (UIBC) in human serum and plasma on Beckman Coulter AU analyzers.",
      "product_quantity": "219 units",
      "reason_for_recall": "The manufacturer has identified that the use of Unsaturated Iron Binding Capacity (UBIC) lot 2690 will produce erroneously high patient results at the low end of the linear range.",
      "recall_initiation_date": "20200818",
      "center_classification_date": "20200930",
      "termination_date": "20210908",
      "report_date": "20201007",
      "code_info": "LOT 2690; Expiration date 10/01/2021"
    }
  ]
}