  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Branchburg",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86364",
      "recalling_firm": "Zeus Scientific, Inc.",
      "address_1": "199 Evans Way",
      "address_2": "N/A",
      "postal_code": "08876-4600",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of CA, CT, IN, NJ, TX and the country of Canada.",
      "recall_number": "Z-3026-2020",
      "product_description": "ZEUS ELISA Parvovirus B19 IgM Test System, UDI 00845533001755 - Product Usage: intended for the qualitative detection of IgM class antibodies to human Parvovirus B19 in human serum including women of childbearing age where there is a suspicion of exposure to human Parvovirus B19.",
      "product_quantity": "74",
      "reason_for_recall": "FDA inspection identified that due to increased positivity (false positives) on the Parvovirus B19 ELISA Test System, the firm assigned a new Correction Factor and new lot number to the remainder of the affected lot, and distributed the reworked product to the customer.",
      "recall_initiation_date": "20190812",
      "center_classification_date": "20200929",
      "termination_date": "20220520",
      "report_date": "20201007",
      "code_info": "Lot 19040014"
    }
  ]
}