  "results": [
    {
      "status": "Terminated",
      "city": "Oro Valley",
      "state": "AZ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80807",
      "recalling_firm": "Ventana Medical Systems Inc",
      "address_1": "1910 E Innovation Park Dr",
      "address_2": "N/A",
      "postal_code": "85755-1962",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "worldwide",
      "recall_number": "Z-3007-2018",
      "product_description": "CONFIRM anti-Estrogen Receptor (ER) (SP1) Rabbit Monoclonal Primary Antibody, Catalog Number 05278414001,  Model 790-4325      Immunohistochemistry (IHC) for in vitro diagnostic use.",
      "product_quantity": "1719 units",
      "reason_for_recall": "Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers.  A false negative IHC result may occur.",
      "recall_initiation_date": "20180802",
      "center_classification_date": "20180914",
      "termination_date": "20211122",
      "report_date": "20180926",
      "code_info": "Lot Numbers:  Y18586, Y24472, UDI: 4015630972234"
    }
  ]
}