  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Crumlin (North)",
      "state": "N/A",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86236",
      "recalling_firm": "Randox Laboratories Ltd.",
      "address_1": "55 Diamond Road",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of WV, and Puerto Rico.",
      "recall_number": "Z-2962-2020",
      "product_description": "Lactate Dehydrogenase L-P, (LDH) (NAO), Catalog Number LD3842 - Product Usage: device intended for the quantitative in vitro determination of Lactate Dehydrogenase (LD) activity in serum and plasma. This product is suitable for use on the RX series instruments which includes the RX daytona and RX imola.",
      "product_quantity": "21",
      "reason_for_recall": "An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional pipette washes can be implemented as described in the technical bulletin.  Interference to the Lactate Dehydrogenase (LOH) reagent would be observed as inconsistencies in Quality Control recovery, which may lead to a delay in running patient samples, or erroneous elevated test results.",
      "recall_initiation_date": "20200811",
      "center_classification_date": "20200911",
      "termination_date": "20201015",
      "report_date": "20200923",
      "code_info": "GTIN: 05055273204124;    Lot Numbers: 480483  495947  523902  536342"
    }
  ]
}