  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-02",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Pittsburgh",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99410",
      "recalling_firm": "Tobii Dynavox Llc",
      "address_1": "2100 Wharton St Ste Ste 400",
      "address_2": "",
      "postal_code": "15203-1691",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2960-2026",
      "product_description": "TD Navio Maxi Speech Generating Device  Model/Catalog Number: 1000799",
      "product_quantity": "",
      "reason_for_recall": "Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.",
      "recall_initiation_date": "20260708",
      "center_classification_date": "20260814",
      "report_date": "20260826",
      "code_info": "UDI-DI 073400747602070  Serial Numbers TDNL1-240904000016 through TDNL1-260617070001.   Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).",
      "more_code_info": ""
    }
  ]
}