      "status": "Ongoing",
      "city": "Danvers",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95129",
      "recalling_firm": "Abiomed, Inc.",
      "address_1": "24 Cherry Hill Dr",
      "address_2": "",
      "postal_code": "01923-2575",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.",
      "recall_number": "Z-2956-2024",
      "product_description": "Impella CP Pump Set, Japan; Product Code: 0048-0034-JP;",
      "product_quantity": "4 units",
      "reason_for_recall": "A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.",
      "recall_initiation_date": "20240805",
      "center_classification_date": "20240904",
      "report_date": "20240911",
      "code_info": "Product Code: 0048-0034-JP; UDI-DI: 00813502011463; Serial Numbers: 324303 323153 328083 328085; Batch Numbers: 2022079852 2022079846 2022083745 2022083747",
      "more_code_info": ""
    }
  ]
}