{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95099",
      "recalling_firm": "Mindray DS USA, Inc. dba Mindray North America",
      "address_1": "800 Macarthur Blvd",
      "address_2": "",
      "postal_code": "07430-2001",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada",
      "recall_number": "Z-2949-2024",
      "product_description": "Sevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system.   Part Number 115-066755-00",
      "product_quantity": "2 units",
      "reason_for_recall": "V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the  patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe  cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention",
      "recall_initiation_date": "20240722",
      "center_classification_date": "20240904",
      "report_date": "20240911",
      "code_info": "UDI-DI: 06936415975784 Serial Number: AH6D28000517, AH6D28000518",
      "more_code_info": ""
    }
  ]
}