      "status": "Ongoing",
      "city": "Norderstedt",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99405",
      "recalling_firm": "Waldemar Link GmbH & Co. KG (Mfg Site)",
      "address_1": "Ost Str. 4-10",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic US distribution to Indiana. OUS distribution to Belarus and Germany.",
      "recall_number": "Z-2934-2026",
      "product_description": "Endo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge  Item Number 15-0027/16",
      "product_quantity": "1 US; 9 OUS",
      "reason_for_recall": "The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.",
      "recall_initiation_date": "20260706",
      "center_classification_date": "20260811",
      "report_date": "20260819",
      "code_info": "UDI-DI 04026575432134 Lot for Implant REF 15-0027/16: 2604069",
      "more_code_info": ""
    }
  ]
}