{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for carboxyhemoglobin (fCOHb) and methemoglobin (fMetHb). A negative interference with fCOHb has the potential to alter the medical assessment of the patient and may withhold necessary follow-up treatment and/or initiate cessation of medication in response to elevated fMetHb levels.",
      "address_2": "Po Box 6101",
      "product_quantity": "10199",
      "code_info": "All units affected",
      "center_classification_date": "20180824",
      "distribution_pattern": "Nationwide distribution. Worldwide foreign distribution.",
      "state": "DE",
      "product_description": "RAPIDPoint 500 Blood Gas Analyzer; 10492730, 1069855, 10696857, 10697306     Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including total hemoglobin.",
      "report_date": "20180905",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-2928-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80516",
      "termination_date": "20200623",
      "recall_initiation_date": "20180716",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}