  "results": [
    {
      "status": "Terminated",
      "city": "Irwindale",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80720",
      "recalling_firm": "Biosense Webster, Inc.",
      "address_1": "15715 Arrow Hwy",
      "address_2": "N/A",
      "postal_code": "91706-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "No U.S. Distribution. International Distribution only.",
      "recall_number": "Z-2878-2018",
      "product_description": "WaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523 and WCR2523.",
      "product_quantity": "25 units",
      "reason_for_recall": "When attempting to recapture the 32-mm device, the tip of the delivery sheath may fold or buckle, resulting in increased retraction force and difficulty or failure to recapture the device.",
      "recall_initiation_date": "20180507",
      "center_classification_date": "20180830",
      "termination_date": "20201114",
      "report_date": "20180905",
      "code_info": "All lots."
    }
  ]
}