      "status": "Ongoing",
      "city": "Bedford",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99115",
      "recalling_firm": "Instrumentation Laboratory",
      "address_1": "180 Hartwell Rd",
      "address_2": "N/A",
      "postal_code": "01730-2443",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2825-2026",
      "product_description": "Brand Name: GEM Premier 5000  Product Name: GEM Premier 5000  Model/Catalog Number: 00024019255   Portable blood analyzer system",
      "product_quantity": "21",
      "reason_for_recall": "Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.",
      "recall_initiation_date": "20260508",
      "center_classification_date": "20260729",
      "report_date": "20260805",
      "code_info": "GEM Premier 5000:  Part number: 00024019255  UDI: 08426950870227  Units distributed in Spain:    GEM Premier 5000 Serial Nos.: 25028719, 18021546, 18021549, 26012036, 18051850, 23054764, 21068342, 25080219, 26010781, 25122012, 25041639, 25120558, 25079944, 25079943, 25028719, 25100485, 25079947, 25100483, 25100484, 21057870 and 26010710.  Affected Accessory:  Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026.  Part No. 00024015859; UDI No. 08426950869634"
    }
  ]
}