  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Horsham",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86023",
      "recalling_firm": "Flower Orthopedics Corporation",
      "address_1": "100 Witmer Rd Ste 280",
      "address_2": "N/A",
      "postal_code": "19044-2647",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "US Nationwide distribution including in the states of AZ, FL, and OH.",
      "recall_number": "Z-2784-2020",
      "product_description": "Calcaneus Plate, Extensile, Large, Left, Catalog Number CPF 103 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of bones of the rearfoot, including the calcaneus.",
      "product_quantity": "25",
      "reason_for_recall": "The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).",
      "recall_initiation_date": "20200708",
      "center_classification_date": "20200812",
      "termination_date": "20201008",
      "report_date": "20200819",
      "code_info": "Lot Number 1909660033; UDI 00840118115667"
    }
  ]
}