{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hartland",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77627",
      "recalling_firm": "Merge Healthcare, Inc.",
      "address_1": "900 Walnut Ridge Dr",
      "address_2": "N/A",
      "postal_code": "53029-8347",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "FL, IL, CO, MI, CA, NY, ND, OH, CT",
      "recall_number": "Z-2754-2017",
      "product_description": "Merge Eye Care PACS Viewer 5.2  Merge Eye Care PACS is a software solution for the display, management, archive, interface and integration of ophthalmic device reports, images and data.   Merge Eye Care PACS is a software solution using databases for patient demographics, server and other systems of intercommunication with hospital systems via HL7 and DICOM to provide to clinicians a single image viewing and management solution of images imported from various contributing devices throughout the clinical environment.  Merge Eye Care PACS provides the ability to review data from any network-connected computer, and is protected by appropriate security login which permits only authorized user access. Symphony uses 256 bit or greater encryption via secure socket layer (SSL) to assure a network environment which is secure.",
      "product_quantity": "18 units",
      "reason_for_recall": "When the user has not set up any user preference on the sorting order to render the study images (OS/OD/etc.) and reports, the ECP may get an incorrect study index. If the user attempts to delete an image, ECP may then read that image as OS when it is, in fact, OD.",
      "recall_initiation_date": "20170613",
      "center_classification_date": "20170720",
      "termination_date": "20200917",
      "report_date": "20170726",
      "code_info": "Version 5.2"
    }
  ]
}