  "results": [
    {
      "status": "Completed",
      "city": "Portage",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "95040",
      "recalling_firm": "Stryker Corporation",
      "address_1": "1941 Stryker Way",
      "address_2": "",
      "postal_code": "49002-9711",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.",
      "recall_number": "Z-2725-2024",
      "product_description": "Stryker iBur 3.0mm Diamond Round, Distal Bend -  intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications  Ref: 8431-012-030D",
      "product_quantity": "2507 units",
      "reason_for_recall": "Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.",
      "recall_initiation_date": "20240724",
      "center_classification_date": "20240826",
      "report_date": "20240904",
      "code_info": "GTIN: 7613327501230 Lot Numbers: 22216017 22230017 22334017 23004017 23004027 23009017 23157017 23208017 23233017 23233027 24003017",
      "more_code_info": ""
    }
  ]
}