  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98869",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Japan, EU, and Latin America.",
      "recall_number": "Z-2721-2026",
      "product_description": "Olympus Glass Body for 194 Ellik Evacuator. Model Number: 190A.",
      "product_quantity": "3,663 units",
      "reason_for_recall": "Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.",
      "recall_initiation_date": "20260505",
      "center_classification_date": "20260710",
      "report_date": "20260722",
      "code_info": "Model Number: 190A. UDI-DI: 00821925003538. All Lot Numbers.",
      "more_code_info": ""
    }
  ]
}