  "results": [
    {
      "status": "Terminated",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69105",
      "recalling_firm": "Horiba Instruments, Inc dba Horiba Medical",
      "address_1": "34 Bunsen",
      "address_2": "N/A",
      "postal_code": "92618-4210",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution.",
      "recall_number": "Z-2692-2014",
      "product_description": "ABX PENTRA Magnesium RTU Reagent, Part No. A11A01646    Product Usage:  ABX PENTRA Magnesium RTU reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of magnesium in human serum and plasma based on a photometric test using xylidyl blue. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemnia (abnormally low plasma levels of magnesium) and hypermagnesemnia (abnormally high plasma levels of magnesium).",
      "product_quantity": "359 units",
      "reason_for_recall": "Horiba Medical is recalling the ABX PENTRA Magnesium RTU Reagent because the on-board stability claims has been updated.  Horiba has updated the ABX PENTRA Magnesium RTU package inserts by removing the 7-day on-board  stability claim and replacing it with a 1-day on-board stability claim.",
      "recall_initiation_date": "20140811",
      "center_classification_date": "20140923",
      "termination_date": "20141125",
      "report_date": "20141001",
      "code_info": "All Lots"
    }
  ]
}