{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Danvers",
      "state": "MA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98893",
      "recalling_firm": "Abiomed, Inc.",
      "address_1": "22 Cherry Hill Dr",
      "address_2": "",
      "postal_code": "01923-2575",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT), Japan (JP), Kuwait (KW), Luxembourg (LU), Malaysia (MY), Mexico (MX), Netherlands (NL), Norway (NO), Panama (PA), Poland (PL), Saudi Arabia (SA), Serbia (RS), Slovenia (SI), South Korea (KR), Spain (ES), Sweden (SE), Switzerland (CH), Taiwan (TW), Thailand (TH), United Arab Emirates (AE), and United Kingdom (GB).",
      "recall_number": "Z-2684-2026",
      "product_description": "ABIOMED Introducer Kit 23 Fr, 11 cm, Pump Set Code 1000441, containing Kit, 23Fr Introducer, 11cm, Sterile, Introducer Product Code 2000342.  heart pump and accessories",
      "product_quantity": "196 units",
      "reason_for_recall": "A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.",
      "recall_initiation_date": "20260611",
      "center_classification_date": "20260710",
      "report_date": "20260722",
      "code_info": "Pump Set Code 1000441; Introducer Product Code 2000342. GTIN: 00813502012910. All current batches in field.",
      "more_code_info": ""
    }
  ]
}