  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85853",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-2662-2020",
      "product_description": "PowerLoc MAX Power Injectable Infusion Port Access Kit; Catalog Numbers 2131910 (UDI 00801741044151), 2131915 (UDI 00801741044168), 2131975 (UDI 00801741044175), 2132010 (UDI 00801741044151), 2132015 (UDI 00801741044199), 2132075 (UDI 00801741044205), 2132210 (UDI 00801741044212), 2132275 (UDI 00801741044236), 2141910 (UDI 00801741044243), 2141915 (UDI 00801741044250), 2141975 (UDI 00801741044267), 2142010 (UDI 00801741044274), 2142015 (UDI 00801741044281), 2142075 (UDI 00801741044298), 2142210 (UDI 00801741044304), 2142215 (UDI 00801741044311), 2142275 (UDI 00801741044328)",
      "product_quantity": "5,296,132 total",
      "reason_for_recall": "Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.",
      "recall_initiation_date": "20200619",
      "center_classification_date": "20200718",
      "termination_date": "20220729",
      "report_date": "20200729",
      "code_info": "Catalog Numbers 2131910 (UDI 00801741044151), 2131915 (UDI 00801741044168), 2131975 (UDI 00801741044175), 2132010 (UDI 00801741044151), 2132015 (UDI 00801741044199), 2132075 (UDI 00801741044205), 2132210 (UDI 00801741044212), 2132275 (UDI 00801741044236), 2141910 (UDI 00801741044243), 2141915 (UDI 00801741044250), 2141975 (UDI 00801741044267), 2142010 (UDI 00801741044274), 2142015 (UDI 00801741044281), 2142075 (UDI 00801741044298), 2142210 (UDI 00801741044304), 2142215 (UDI 00801741044311), 2142275 (UDI 00801741044328).      Lot Numbers:  REDW3798  REDY2866  REDZ1719  REEN3137  REEP0951  REEQ2684  REEQ4092  REER1229  REEQ0664  REEQ1649  REER0459  REDZ1818  REEP3049  REDW1092  REDX3754  REDY1682  REDY2971  REDY3530  REDZ1070  REEN3271  REEN4480  REEP0992  REEP3957  REEQ1647  REEQ2724  REEQ4093  REER1266  REDY3719  REEN1270  REEP0952  REEQ0684  REEQ2676  REEQ4054  REER0460  REDT2879  REDU2256  REDU3371  REDU4487  REDV1088  REDV2972  REDV3470  REDW0402  REDW1087  REDX1518  REDY1804  REDY2867  REDY3531  REDZ1069  REEN1104  REEN4481  REEP0953  REEP3958  REEQ0665  REEQ1650  REER0530  REEN2139  REEP4053  REER0461  REDY3720  REEQ4094  REDW3682  REDY0827  REDZ1071  REEP0994  REER0539  REEN4551  REEP0954  REEQ1648  REDW1197  REDY2972  REDZ3130  REEP0995  REEP2909  REDW0546  REDX1519  REDX4930  REDY1683  REDY2868  REDY3532  REDZ1064  REDZ1819  REEN1046  REEN2146  REEP0955  REEP3052  REEP3959  REER0462  REDZ1721  REEQ2678  REDV1236  REDW0407  REDW0550  REDX4852  REDY0773  REDY0828  REDY1684  REDY2973  REDZ1063  REDZ1072  REDZ1820  REEN1271  REEN3273  REEP0956  REEP0996  REEP1728  REEP3053  REEP4054  REEQ1664  REEQ2720  REER0540  REDW3683  REDY3721  REEN4552  REER1261  REEN3155  REDW0403  REDY1685  REDZ1722  REEP0997  REEQ0666",
      "more_code_info": ""
    }
  ]
}