  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Salt Lake City",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99096",
      "recalling_firm": "Bard Access Systems, Inc.",
      "address_1": "605 N 5600 W",
      "address_2": "N/A",
      "postal_code": "84116-3738",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.",
      "recall_number": "Z-2661-2026",
      "product_description": "5855240\tPower-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen  UDI-DI Code: 00801741107177    5855300\tPower-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen  UDI-DI Code: 00801741107184    5865120\tPower-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen  UDI-DI Code: 00801741107191    5865200\tPower-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen  UDI_DI Code: 00801741107214    5865240\tPower-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen  UDI-DI Code: 00801741107221      The Power-Trialysis Slim-Cath Short-Term Dialysis Catheter, with a third internal lumen for intravenous therapy, power injection of contrast media, and central venous pressure monitoring, is indicated for use in attaining short-term (less than 30 days) vascular access for hemodialysis, hemoperfusion, and apheresis treatments. The catheter is intended to be inserted in the jugular, femoral, or subclavian vein as required. The maximum recommended infusion rate is 5 mL/sec for power injection of contrast media.",
      "product_quantity": "915",
      "reason_for_recall": "Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.",
      "recall_initiation_date": "20260521",
      "center_classification_date": "20260702",
      "report_date": "20260708",
      "code_info": "5855240 Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107177 Lot Number: REKY2412  5855300 Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107184 Lot Number: REKY1364  5865120 Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107191 Lot Number: RELN0881  5865200 Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI_DI Code: 00801741107214 Lot Number: REKY1381  5865240 Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107221 Lot Number: REJZ0927"
    }
  ]
}