  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Franklin Lakes",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85853",
      "recalling_firm": "Becton Dickinson & Company",
      "address_1": "1 Becton Dr",
      "address_2": "N/A",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution",
      "recall_number": "Z-2660-2020",
      "product_description": "PowerLoc MAX Power Injectable Infusion Set AllPoints Port Access System; Catalog Numbers AN132075 (UDI 00801741118968), AN142010 (UDI 00801741119231), AN142010G (UDI 00801741119323), AN142075 (UDI 00801741119224)",
      "product_quantity": "5,296,132 total",
      "reason_for_recall": "Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.",
      "recall_initiation_date": "20200619",
      "center_classification_date": "20200718",
      "termination_date": "20220729",
      "report_date": "20200729",
      "code_info": "Catalog Numbers AN132075 (UDI 00801741118968), AN142010 (UDI 00801741119231), AN142010G (UDI 00801741119323), AN142075 (UDI 00801741119224).    Lot Numbers:  REDT4257  REDV3617  REDY1815  REDY2789  REDZ0385  REEN2192  REEN2782  REDW2682  REDX1471  REDX3857  REDY0765  REDY1593  REDY2943  REDZ0676  REDZ1903  REDZ3077  REEN1222  REEN3292  REEP0666  REEP4137  REER0612  REDW1021  REDX1360  REDX4728  REDY1816  REDY2935  REDY3765  REDZ1825  REDZ3146  REEN1935  REEP0773  REEP3544  REEP4066  REER0579  REER1103  REDV2172  REDV3683  REDW0467  REDW0521  REDW1023  REDW1157  REDX1361  REDX1490  REDY0815  REDY1594  REDY2790  REDY3544  REDZ0110  REDZ0386  REDZ0677  REDZ1904  REDZ3078  REEN1072  REEN2186  REEN3075  REEN3310  REEP0663  REEP1401  REEP4138  REEQ0611  REEQ1246  REEQ2436  REEQ3589  REER0576  REER1046  REER2161  Additional lot added 10/16/20: REEU0314",
      "more_code_info": ""
    }
  ]
}