  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69112",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Canada, Chile, Mexico, Panama, and Thailand.    .",
      "recall_number": "Z-2651-2014",
      "product_description": "Coulter DxH Diluent, Catalog No. 628017    Blood cell analysis comprises diluting a whole-blood sample with a solution that functions as a diluent. The diluent provides the ability to analyze portions of the diluted blood sample for different blood cell types and the measurement of hemoglobin.",
      "product_quantity": "46,050 units total (40,078 units in US)",
      "reason_for_recall": "Beckman Coulter is recalling Coulter DxH Diluent because it may be associated with elevated platelet (PLT) background counts between 3-7 x 10(3) cells/uL. This will result in failed Daily Checks on the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems.",
      "recall_initiation_date": "20140825",
      "center_classification_date": "20140916",
      "termination_date": "20150225",
      "report_date": "20140924",
      "code_info": "Lot No. 3510470, 3510480, 3510490, 3510500, 3510510, 3510520, 3510530, 3510540, 3510550, 3510570, 3510580, 3510590, 3510600, 3510610, 3510620, 3510630, 3510690.",
      "more_code_info": ""
    }
  ]
}