      "status": "Ongoing",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92967",
      "recalling_firm": "CareFusion 303, Inc.",
      "address_1": "10020 Pacific Mesa Blvd",
      "address_2": "N/A",
      "postal_code": "92121-4386",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: FL, MI, KS",
      "recall_number": "Z-2646-2023",
      "product_description": "BD Pyxis MedBank MedPass Software, REF: 139088-01",
      "product_quantity": "37",
      "reason_for_recall": "After dispensing a timed dose medication for a patient, the automated dispensing cabinet software may continue to show the already dispensed medication dose is still available for dispense, which could lead to duplicate medication administration.",
      "recall_initiation_date": "20230818",
      "center_classification_date": "20230928",
      "report_date": "20231004",
      "code_info": "UDI-DI: 10885403513817, Software versions prior to but not including v1.2.3.23"
    }
  ]
}