  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Best",
      "state": "N/A",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97501",
      "recalling_firm": "PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.",
      "address_1": "Veenpluis 4-6",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.",
      "recall_number": "Z-2624-2025",
      "product_description": "Allura Xper FD10/10 Product Codes: (1)722011, (2) 722027, (3) 722005; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;",
      "product_quantity": "286 units (67 US, 219 OUS)",
      "reason_for_recall": "Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.",
      "recall_initiation_date": "20250903",
      "center_classification_date": "20250925",
      "report_date": "20251001",
      "code_info": "Product Codes: (1)722011, (2) 722027, (3) 722005; UDI-DI: (1) 00884838059047, (2) 00884838054196, (3) N/A; Serial numbers: (1) 157, 268, 95, 107, 59, 319, 49, 170, 173, 101, 84, 155, 133, 234, 345, 332, 102, 270, 177, 132, 325, 15, 87, 255, 171, 362, 168, 275, 344, 324, 163, 103, 215, 328, 286, 81, 293, 294, 224, 117, 334, 338, 146, 145, 266, 21, 202, 165, 367, 372, 340, 318, 98, 211, 158, 271, 204, 246, 206, 321, 295, 104, 217, 265, 261, 125, 282, 216, 333, 329, 230, 198, 244, 292, 180, 183, 314, 337, 341, 376, 377, 378, 259, 347, 287, 99, 233, 156, 71, 77, 188, 223, 238, 153, 129, 278, 118, 186, 108, 89, 258, 310, 274, 139, 141, 86, 300, 88, 309, 227, 220, 242, 166, 299, 273, 190, 189, 260, 280, 74, 75, 251, 228, 100, 199, 194, 91, 116, 112, 114, 43, 313,           (2) 123, 133, 134, 183, 106, 169, 107, 42, 101, 209, 153, 40, 156, 112, 1, 104, 71, 175, 116, 174, 29, 50, 56, 152, 43, 49, 202, 178, 127, 39, 27, 64, 78, 99, 6, 86, 7, 145, 102, 44, 142, 30, 79, 38, 85, 181, 144, 177, 165, 201, 3, 68, 51, 21, 23, 24, 90, 171, 45, 122, 33, 137, 163, 150, 96, 155, 20, 166, 147, 139, 91, 136, 193, 69, 65, 168, 206, 158, 66, 75, 62, 138, 121, 18, 31, 131, 195, 192, 125, 148, 149, 46, 47, 135, 159, 100, 103, 110, 57, 274, 61, 36, 151, 300, 167, 60, 81, 126, 58, 184, 5, 188, 94, 154, 172, 35, 76, 95, 187, 170, 146, 162, 82, 117, 48, 205, 182, 132,          (3) 212, 435, 128, 323, 269, 342, 531, 403, 287, 457, 373, 332, 374, 428, 75, 499, 511, 473, TPE, 333, 237, 338, SN545, 159, 100071298SN408, 96;",
      "more_code_info": ""
    }
  ]
}