  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Addison",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85517",
      "recalling_firm": "Skeletal Kinetics, Llc",
      "address_1": "3885 Arapaho Rd",
      "address_2": "N/A",
      "postal_code": "75001-4314",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of  AL, CA, CO, FL, IL, MO, MS, OH, OK, OR, TN, TX, WI and the countries of Australia, Canada, South Africa, Switzerland.",
      "recall_number": "Z-2623-2020",
      "product_description": "OSTEOVATIONEX, INJECT 3CC.  Product Number: 390-0003, UDI: 813845020016 - Product Usage: Callos is indicated to fill bony voids or gaps of the skeletal system (i.e. extremities, spine, pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. Callos Bone Void Filler is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.",
      "product_quantity": "101",
      "reason_for_recall": "Bone void filler kits may not maintain its setting characteristics for the labeled shelf-life duration. Use may increase the required surgical and anesthesia time for the patient or possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.",
      "recall_initiation_date": "20200406",
      "center_classification_date": "20200715",
      "termination_date": "20210224",
      "report_date": "20200722",
      "code_info": "Lot Numbers:  17100402,  1127942,  1130302,  1135740,  1142070,  1142943."
    }
  ]
}