  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Flower Mound",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97285",
      "recalling_firm": "Stryker Communications",
      "address_1": "571 Silveron Blvd",
      "address_2": "N/A",
      "postal_code": "75028",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of TX, IL, NY, UT, OR, AZ, VT, NY, IL, MN.",
      "recall_number": "Z-2620-2025",
      "product_description": "Chromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P60034",
      "product_quantity": "40",
      "reason_for_recall": "Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur.",
      "recall_initiation_date": "20250715",
      "center_classification_date": "20250924",
      "report_date": "20251001",
      "code_info": "UDI-DI: 07613327296167 and 07613327651317 REF/Serial Numbers: CH00000001/7000010116, 7000010117, 7000010118, 7000010119, 7000010120, 7000010121, 7000010122, 7000010123, 7000010124, 7000010125, 7000010141, 7000010142, 7000010177, 7000010178, 7000010148, 7000010149, 7000010157, 7000010137, 7000010175, 7000010176, 7000010145, 7000010146, 7000010147, 7000010309, 7000010305, 7000010307, 7000010296, 7000010143, 7000010150, 7000010151, 7000010155.  P60034/1000022, 1000023, 1000024, 1000016, 1000017, 1000018, 1000019, 1000020, 1000021",
      "more_code_info": ""
    }
  ]
}