  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99209",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "N/A",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.",
      "recall_number": "Z-2613-2026",
      "product_description": "Medline Convenience kits containing recalled Swan-Ganz Catheters    CABG A CDS\t  \tCDS982269K  \tCDS982269L  \tCDS982269M  \tCDS982269N  \t  CARDIAC MAJOR\t  \tDYNJ901328D  \tDYNJ901328F  \t  HEART A\t  \tDYNJ903465F  \tDYNJ903465G  \t  HEART TRAY-VASCULAR\t  \tDYNJ907857  \tDYNJ907857A  \tDYNJ907857B  \t  KIT UNIVERSITY CABG PK\t  \tDYNJ905857A  \t  OPEN HEART\t  \tCDS981161AA  \tCDS981161AB  \tDYNJ905437A  \tDYNJ909090  \tDYNJ909090C  \tDYNJ9855016R  \tDYNJ9855016S  \t  OPEN HEART ACCESSORY CDS\t  \tCDS984289J  \t  OPEN HEART ANESTHESIA PACK\t  \tCDS983849D  \tCDS983849F  \tCDS983849G  \t  OPEN HEART B CDS\t  \tDYNJ900377P  \t  OPEN HEART BASIC B\t  \tDYNJ904311C  \t  OPEN HEART CDS\t  \tCDS983513F  \t  OPEN HEART CDS QVH\t  \tCDS982523K  \tCDS982523L  \tCDS982523N  \t  OPEN HEART LINE SET UP\t  \tDYNJ908179  \tDYNJ908179A  \t  OPEN HEART SUPPLEMENT\t  \tDYNJ909116  \tDYNJ909116A  \tDYNJ909116C  \tDYNJ909116D  \t  SILVER CROSS OPEN HEART\t  \tDYNJ906102J  \tDYNJ906102K  \tDYNJ906102L  \t  \t  SM OPEN HEART PACK A&B\t  \tDYNJ904749G  \tDYNJ904749J  \t  VPH OPEN HEART\t  \tDYNJ902664J  \tDYNJ902664K  \tDYNJ902664L",
      "product_quantity": "9,272 kits",
      "reason_for_recall": "Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.",
      "recall_initiation_date": "20260528",
      "center_classification_date": "20260626",
      "report_date": "20260708",
      "code_info": "CDS982269K    UDI-DI 10193489906844    Lots 21HBN691   21JBQ518   21LBD444  CDS982269L    UDI-DI 10195327018016    Lots 21LBP501   22ABR514   22BBE936   22FBV989  CDS982269M    UDI-DI 10195327192808    Lots 22JBE028   22LBF124   23ABP909   23CBO580  CDS982269N    UDI-DI 10195327352998    Lots 23GBC969   23HBS959   23KBW067   23LBF510   24CBO691   24DBR250   24EBU619   24GBF288   25ABM339   25CBJ666   25HBH657   25IBM477   26BBG451    DYNJ901328D    UDI-DI 10193489854923    Lots 21CBQ397   21FBJ935   21HBR220   21JBV897   21JBV898   22ABG611  DYNJ901328F    UDI-DI 10195327044572    Lots 22CBX749   22FBX103   23ABC195   23EBG860   23EBL637   23EBS968   23FBM688   23GBS043   24EBL355   24FBJ301   24IBS567   24JBD767   24JBU518   24KBL806   24LBF815   25DBF590   25FBQ740   25GBH267   25JBQ634    DYNJ903465F    UDI-DI 10193489918397    Lots 21FDB453   21HDB016   21IDB927   21KDA273   21LDA283   21LDB750   22CDA262  DYNJ903465G    UDI-DI 10195327175559    Lots 22FDA593   22IDA677   22KDA077   23ADB246    DYNJ907857    UDI-DI 10193489909364    Lots 21GBJ159   21IBX945   21LBJ320   22ABH627  DYNJ907857A    UDI-DI 10195327133160    Lots 22HBK592   22KBM436  DYNJ907857B    UDI-DI 10195327256869    Lots 22KBN208   23ABM669   23IBE277    DYNJ905857A    UDI-DI 10193489946680    Lots 22CBW896   22DBR508   22GBG113   22GBW887   22IBU528   22JBA758   22KBJ132   23ABP277   23ABP278   23ABS778   23DBD840   23EBJ932   23EBR134   23FBM591   23FBS624   23HBU190   24FBG082   24GBE581   24HBM636   24IBS825   24JBV903   24KBU708   24LBU110   25BBG391   25CBJ747   25DBP832    CDS981161AA    UDI-DI 10193489303407    Lots 21DBH590   21DBU485  CDS981161AB    UDI-DI 10193489817461    Lots 21DBT869   21FBD406   21FBK899   21HBT158   21IBJ277   21IBW875   22BBA363   22DBH285   22DBT988   22HBO231   22HBT037   22JBT785  DYNJ905437A    UDI-DI 10193489571905    Lots 21ABH889   21DBE110   21EBH610  DYNJ909090    UDI-DI 10195327268350    Lot 23EBD437  DYNJ909090C    UDI-DI 10198459084768    Lots 24HBR121   24KBT550  DYNJ9855016R    UDI-DI 10193489481730    Lot 21DKA468  DYNJ9855016S    UDI-DI 10193489904659    Lots 21HBC511   21HBC512   22GBZ353   22JBA413   22OBI292   22OBL745   23BBA924   23BBN806   23CBM193    CDS984289J    UDI-DI 10193489263046    Lots 21BKA389   21DKA676   21FLA954   21GLA057   21GLA312   21GLB087   21HLA680   21ILA628   21KLB205    CDS983849D    UDI-DI 10193489278231    Lots 21ABW164   21BBR951   21CBU594   21DBV510   21EBA170   21FBS444   21GBM142   21HBK335   21JBS153   21JBX783   21LBF083   22BBY971   22CBU521   22EBK057   22EBV010   22GBJ239   22GBZ470   22OBM677  CDS983849F    UDI-DI 10195327228316    Lots 22JBN998   22KBB265   22LBT079   23ABN910   23ABO447   23DBG442   23EBP559   23FBP234   23FBV862   23GBN607   23HBA321   23IBM362   23JBQ839   24ABE486   24BBC426  CDS983849G    UDI-DI 10195327650148    Lots 24EBA261   24FBD116   24GBH789   24HBQ756   24HBV600   24IBV160   24LBB436   25ABD357   25ABL507   25ABO737   25BBM406   25CBL909   25EBN768   25GBC295   25HBJ455   25JBJ254   25LBJ730   25LBQ452   26ABV988   26CBR016   26EBG732    DYNJ900377P    UDI-DI 10193489818710    Lots 21ABS254   21DBG173   21EBM044   21FBJ060   21GBC753    DYNJ904311C    UDI-DI 10193489427141    Lots 21CBD905   21DBT916    CDS983513F    UDI-DI 10193489843729    Lots 21FBF442   21GBV619   21KBR342   22ABG599   22BBX671   22GBI891   22HBI664   22KBI289   23BBE439    CDS982523K    UDI-DI 10193489372199    Lots 21AKB056   21CKA531   21DKA456   21EKA149   21FKA087   21GKA019   21HKA157   21HKA477   22ALA702  CDS982523L    UDI-DI 10195327111151    Lots 22CLB415   22DLA940   22FLA279   22HLA254   22JLA168   22JLA569   22LLA984   23ALA240   23ALA433   23BLA324   23CLA025   23DLA469   23DLB096   23HLB221   23ILA547   23JLB156  CDS982523N    UDI-DI 10198459094187    Lots 24JLA242   25ALA328   25BLA052   25CLA357   25ELB103   25GLA805   25HLA714   25JLA015   25JLA997   25LLA549   26ALA715   26BLA627    DYNJ908179    UDI-DI 10193489999082    Lot 21KBU602  DYNJ908179A    UDI-DI 10195327222574    Lots 22IBI258   23ABC799    DYNJ909116    UDI-DI 10195327275419    Lot 22LBF704  DYNJ909116A    UDI-DI 10195327326869    Lot 23BBA820  DYNJ909116C    UDI-DI 10195327412234    Lots 23EBI448   23GBT021   23HBO369   23LBE724   24ABE871   24ABU718   24CBD865   24CBQ303   24DBT408  DYNJ909116D    UDI-DI 10198459055355    Lots 24GBV993   24HBL885   24IBD459   24JBB975   24KBN648   25ABB011   25BBC914   25EBD155   25FBM604   25GBP604   25HBE896   25JBM466   25KBS952   25LBK344   26CBI558   26CBV164   26DBF078    DYNJ906102J    UDI-DI 10195327560485    Lots 24DBU434   24HBL681  DYNJ906102K    UDI-DI 10198459079733    Lots 24IBB695   24IBF531   24JBC419   24KBA747   24KBJ996  DYNJ906102L    UDI-DI 10198459242564    Lot 25GBE498    DYNJ904749G    UDI-DI 10195327368876    Lot 23GBJ246  DYNJ904749J    UDI-DI 10195327511142    Lots 24KBJ796   24LBP688   25ABE593   25BBG734   25EBU720   25HBI847   25IBQ209   25IBU994   25KBC533   26ABM369   26DBI408    DYNJ902664J    UDI-DI 10193489827460    Lots 21CKA218   21DKA466   21EKA130   21GLA739   21HLA316   21ILA068   21ILB049   21LLA925   22BLA932  DYNJ902664K    UDI-DI 10195327102524    Lots 22CLA698   22DLA226   22GLA416   22ILB123   22KLA463   23ALA437   23GLA545   23HLB280   23ILA818  DYNJ902664L    UDI-DI 10195327537883    Lots 24ELA957   24GLA394   24ILA830   25CLA878"
    }
  ]
}