  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Vernon Hills",
      "address_1": "353 Corporate Woods Pkwy",
      "reason_for_recall": "The distal end of the Conical portio adapter (acorn) is composed of a small cap (head) that partially sits inside a hollow shaft (tube). These two parts are welded together to create one piece. In the  impacted batches of product, the weld was improperly performed. As a result, the two pieces are not properly joined. This potential failure can result in the separation of the two pieces. The possible separation of these two components presents a potential risk to patients.",
      "address_2": "",
      "product_quantity": "Product code 8378.90- 94 pieces & Product code 8378.91 - 32 pieces",
      "code_info": "Product number 8378.90 with Batch number 51005921 and Product number 8378.91 with Batch number 51005922",
      "center_classification_date": "20160819",
      "distribution_pattern": "U.S. Consignees: AK, AL, AR, AZ, CA, CT, DC, FL, GA, IA, IL, KY, MA, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OR, TX, UT and VA.",
      "state": "IL",
      "product_description": "Conical portio adapter (acorn) small & large are used for examination and diagnosis in conjunction with endoscopic accessories in gynecological laparoscopy, to determine patency of the fallopian tubes.",
      "report_date": "20160831",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Richard Wolf Medical Instruments Corp.",
      "recall_number": "Z-2608-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74656",
      "termination_date": "20171130",
      "more_code_info": "",
      "recall_initiation_date": "20160519",
      "postal_code": "60061-3110",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}