  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Austin",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92969",
      "recalling_firm": "Encore Medical, LP",
      "address_1": "9800 Metric Blvd",
      "address_2": "",
      "postal_code": "78758-5445",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: AL, AZ, CA, CO, FL, IL, IN, KS, KY, LA, MA, MI, MN, MO, NC, NJ, NY, OH, PA, PR, RI, SC, TX, UT, VA, WA.  OUS: Sweden, Australia, France, UK, Colombia, Germany",
      "recall_number": "Z-2604-2023",
      "product_description": "DJO Surgical: Delta Ceramic Femoral Head, 44 mm + 4.0 mm, REF: 400-03-443",
      "product_quantity": "10",
      "reason_for_recall": "Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery",
      "recall_initiation_date": "20230804",
      "center_classification_date": "20230919",
      "report_date": "20230927",
      "code_info": "UDI-DI/Lot/Expiration: 00888912076210/878B1171/May 2, 2027",
      "more_code_info": ""
    }
  ]
}